In lipid analysis we usually have three steps of workflow generating a series of raw data -> interpreted data -> integrated data.
GC-FID and GC-MS (scan and SIM) data is generated by masshunter software.
For example, there is a sequence of measurements.

Each measurement will generate a folder containing a series of raw data (method acquisition file and data files).

GC-C-IRMS data is generated by Qtegra.
Each sequence produces a single container file (.imexp) that contains all the measurement data. It is not possible to extract individual measurements from this file. The .imexp format can only be opened with Qtegra® software.

After initial checks and corrections in Qtegra®, the relevant data is exported to .xlsx format for calibration and for general accessibility.

It's also possible to export all chromatogram data points, allowing graphs to be reconstructed in Excel (.xlsx) or ohter programs as well.

GC-FID/MS raw data is interpreted by masshunter software. We’ll identify the compounds (peaks) in each chromatogram and summarize the compound information in an Excel table (interpreted data).


The interpreted data will be further integrated by statistics or modelling (integrated data).
Data reliability/limitation of method
Quantification is a problem for statistics. Every instrument (GC-MS and GC-C-IRMS) generates data under its specific quantitative scale. We are doing qualitative and semi-quantitative analysis for GC-MS. The interpreted data is usually (1 vs 0) present/absent or the ratio/percentage of two/several compounds. But other analyses like GC-C-IRMS are quantitative, which generate accurate concentration data with exact quantified numbers (for example, 25.68 ‰ of carbon isotopes). It’s hard to do statistics when we integrate qualitative GC-MS data with other quantitative data.